23/07/2020

Single-step calibration method for nano indentation testing machines

See our publication: M. Galetto, G. Genta, G. Maculotti. Single-step calibration method for nano indentation testing machines application/pdf (945.76 kB). Published on: CIRP ANNALS - MANUFACTURING TECHNOLOGY. DOI: 10.1016/j.cirp.2020.03.015

Nano indentation is an effective method for materials mechanical characterisation at grain scale. Literature underlined relevance of testing machine calibration to measurement uncertainty of the mechanical characterisation. ISO 14577-2 defines multi-step iterative methods for calibrating frame compliance and indenter area function that do not require high-resolution microscopes. Previous research demonstrated that standard’s recommendations are unsatisfactory and result in high calibration uncertainty. This work defines an improved calibration method based on a single-step procedure that achieves, as proved by experimental tests, definite advantages in terms of implementation and measurement uncertainty of both calibration and mechanical characterisation with respect to current procedures.

Published on: 23/07/2020